Configuring a Spacelabs system for the Vocera Platform

These settings enable direct communication between Spacelabs CEI (Clinical Event Interface) and the Vocera Platform.

Connect the Spacelabs ICS (Intesys Clinical Suite) server, the Spacelabs qube patient monitor, and a standard computer to the network hub or switch. Power on the server and log in with the supplied credentials.

Note: For more information on configuring the Spacelabs Clinical Event Interface, see Section 13 of the Intesys Clinical Suite G2 Administration Manual.
  1. Launch the ICS Administration tool from the Windows desktop or Program Files menu, and navigate to the CEI Server tab.
  2. Complete the configuration fields as described in the table.
    Configuration Field Description
    CEI Server Listening Port Change the default port (6868) to the CEI server port number entered in the Configuring a Vocera Spacelabs CEI Adapter. The same port must be used by the CEI server and the Vocera Spacelabs CEI Adapter to enable communication.
    Alarm Notification Mode Select the alarm processing mode for the CEI server. Vocera recommends Alarm Text Only, to send alarms while omitting waveforms. Do not select Alarm Notification Disabled, as no alarms will be sent to Vocera Platform.
    Vital Signs Update Select the Update Disabled checkbox. Currently, Vocera Platform does not process vital signs updates.